DOCUMENTATION

Connecting instruments

Analysers send results to Maisha LabFlow through a small connector agent on a computer in your laboratory. Interfaced results go through the same verification and release steps as results entered by hand.

Reviewed October 2026 · Maintained by the Maisha Health product team

Before you start

Tell us which analysers you use. Interfaces differ by model and firmware, so we confirm each connection with you before you rely on it, and we set up the first connections during onboarding.

1. Register a connector

Under Equipment → Analyser interfacing → Connector agents, register a connector for the analyser and choose its branch. A device token is shown once; copy it into the connector agent's configuration. The status card shows the connector online once it has checked in within the last five minutes.

2. Message formats

The connector sends results over HTTPS. It accepts ASTM E1394 and HL7 v2 result messages, and structured JSON. Where an analyser supports host query, it can scan a specimen barcode and receive the tests ordered for it. Results can also be imported from a file on the Results page.

3. Map test codes

Each analyser uses its own codes. In Test Catalog, open a test and add its analyser codes under Analyzer mappings. Codes that are unmapped, ambiguous or not ordered for that specimen are reported back for review rather than written to a patient's record.

4. Review and release

Interfaced results arrive as pending and wait for verification and release like any other result. Your laboratory can choose to hold results that arrive without a recorded operator until someone adds a note. Reagent lots reported by the analyser are matched against your stock, and mismatches are listed for review.

Run log

Equipment → Analyser interfacing → Run log shows each message received and what happened to it, so you can trace any result back to the analyser run.

Questions? Talk to us →